Skip to main content

Table 5 Disinfection effect on stainless steel carriers surface by LPM lamp (temperature of the carriers was 20 °C; bacterial suspensions were prepared in TSB)

From: Study on disinfection effect of a 222-nm UVC excimer lamp on object surface

Diluent for bacterial suspension; treatment for stainless steel carriers

Mean log10 CFU in positive control group (range)

Mean KL ± SD for different irradiation time (min)

10

20

30

TSB; not dried

6.17 (6.12–6.23)

Staphylococcus aureus

3.31 ± 0.02

5.17 ± 0.25

5.77 ± 0.05

6.12 (6.08–6.16)

Escherichia coli

3.33 ± 0.03

4.86 ± 0.20

5.62 ± 0.14

6.11 (6.06–6.15)

Pseudomonas aeruginosa

3.39 ± 0.01

5.10 ± 0.25

5.51 ± 0.13

TSB; dried

6.37 (6.28–6.46)

Staphylococcus aureus

2.92 ± 0.03

4.29 ± 0.03

4.62 ± 0.06

6.33 (6.23–6.41)

Escherichia coli

2.95 ± 0.02

4.33 ± 0.01

4.82 ± 0.03

6.29 (6.16–6.40)

Pseudomonas aeruginosa

2.95 ± 0.02

4.31 ± 0.04

4.87 ± 0.26

  1. SD standard deviation